For Tubidity Testing

Designed to provide the accuracy demanded by today’s scientist, the Micro100 Laboratory Turbidimeter is ready to meet the challenge. Features like the Auto Alert Calibration Prompt and the Quick Connection Lamp Module, allow for simple calibration and low maintenance. With a resolution of 0.01 NTU at low turbidity readings and an extended range to 1,000 NTU, the Micro100 is an ideal and affordable tool for research as well as routine analytical measurements.

Available Models

Technical Specifications 

Principal of operationNephelometric non-ratio
RangeAuto ranging from 0 – 1000 NTU
Accuracy± 2% of reading plus 0.01 NTU
Resolution0.01 NTU (0 – 9.99 NTU)0.1 NTU (10 – 99.9 NTU)1 NTU (100 – 1000 NTU)
Response TimeLess than 6 seconds
Sample size30 ml (27 ml minimum)
Light sourceQuick connect tungsten filament or infrared
Temperature32° – 122°F (0° – 50°C)
OutputsRS-232 serial port
Dimensions10.75 x 10 x 3.75 inches (27.3 x 25.4 x 9.5 cm)
Weight2.9 lbs. (1.32 kg)
Safety RatingUL, CSA, CE
Weight7.0 lbs. (3.18 kg)
CertificationUL, CSA, CE, USEPA, ISO7027

Micro100 IR Laboratory Turbidimeter, 0 – 1000 NTU, Universal Power 100-240V 50-60 Hz Includes a complete set of calibration standards (0.02, 10 and 1000 NTU), two measuring cuvettes with light shield cap, operator’s manual and a universal power adapter that operates at any voltage between 100-240VAC.

Technical Specifications 

Principal of operationNephelometric
Range0 – 10,000 NTU0 – 2,450 EBC0 – 10,000 FNU0 – 10,000 FAU
Ratio (color correction)Full time ON or OFF
Accuracy± 2% of reading plus 0.01 NTU (0 to 1000 NTU)± 5% of reading (1000 to 4000 NTU)± 10% of reading (4000 to 10,000 NTU)
Resolution0.0001 NTU on lowest range
Response timeLess than 6 seconds
Sample size30 ml
Light sourceQuick connect tungsten filament or infrared
Operating temperature32° – 122°F (0° – 50°C)
Air purgeConnection for external dry air supply
OutputsRS-232 serial port (bidirectional)
Shipping dimensions14 x 12 x 12 inches (35.5 x 30.4 x 30.5 cm)

Micro100 IR Laboratory Turbidimeter, 0 – 1000 NTU, Universal Power 100-240V 50-60 Hz Includes a complete set of calibration standards (0.02, 10 and 1000 NTU), two measuring cuvettes with light shield cap, operator’s manual and a universal power adapter that operates at any voltage between 100-240VAC.

Key Features

  • Auto ranging 0 – 1000 NTU: Instrument senses turbidity level of sample and automatically adjusts to the appropriate measurement
  • Auto alert calibration prompt: Automatically prompts the operator when calibration of the instrument is needed
  • Simple calibration procedures: Calibration initiated with the push of a button ensures accurate readings
  • RS-232 output: Date, time, and NTU reading can be captured with a serial printer or data recorder

Key Industries

  • Aerospace
  • Agriculture
  • Automotive
  • Biopharmaceutical and Biotechnology
  • Chemical
  • Construction
  • Cosmetics & Personal Care
  • Defense
  • Die & Mould
  • Education & Academics
  • Electronics
  • Energy
  • Environmental
  • Food & Beverage
  • Luxury
  • Machinery & Manufacturing
  • Medical
  • Mining & Minerals
  • Oil & Gas
  • Optical
  • Paper & Wood
  • Pharmaceutical
  • Polymers, Plastics and Rubbers
  • Powders & Pigments / Coating
  • Railway
  • Semiconductor, Solar & Electronics
  • Service
  • Textile
  • Wholesale & Retail

Brand

HF Scientific

HF Scientific is a quality innovator of instrumentation focused on the measurement of cleanwater. They offer laboratory and on-line instruments to monitor important parameters in water and wastewater treatment plants.

Your download – just a few steps away

Please enter your details here:

test

Contact us